Aeroneb Go Instruction Manual

Browse online or download Instruction Manual for Unknown Aeroneb Go. Aeroneb Go Instruction manual User Manual

  • Download
  • Add to my manuals
  • Print

Summary of Contents

Page 2

6CLASSIFICATIONS AND MARKINGSThe symbols that appear on the Aeroneb® Go Nebulizer and its accessories are described in the table below:Symbol MeaningA

Page 3 - Table of Contents

7AERONEB® GO SYSTEM PARTS:Medication Cup CapMouthpieceNebulizer bodyCableBaseControl Module AC/DC Adapter** Available with complete Aeroneb® Go System

Page 4

8ASSEMBLY/FUNCTIONAL TEST1. Connect the Medication Cup to the Nebulizer Body by lining up the notches on the medication cup and nebulizer body.NOTE:

Page 5

9HOW TO OPERATE THE AERONEB® GO NEBULIZER3 x AA LR61. Turn the Aeroneb® Go Control Module upside down. Slide the battery cover in the direction of th

Page 6

10 5. Alternatively, you can power the device by using the AC/DC Adapter if provided (refer page 7). Plug the AC/DC adapter into a standard AC electr

Page 7

11CLEANING AND STORING YOUR AERONEB® GOAFTER EVERY USE:Disconnect the control module before you take the nebulizer apart to rinse it.1. Rinse the Neb

Page 8

12OPTIONAL DISINFECTION:If you would like to disinfect your Aeroneb® Go Nebulizer, you may choose one of the following two options; Boiling or Vinegar

Page 9

13OPTIONAL DISINFECTION2. VINEGAR SOLUTION:i. Rinse the Nebulizer Unit (Medication Cup, Nebulizer Body, Base and Mouthpiece or Mask Adapter) under r

Page 10

14TROUBLESHOOTING GUIDEIf you have difculty operating your Aeroneb® Go, check the following points:Problem Check RemedyNo visible aerosol when using

Page 11 - Go System

15Problem Check RemedyCheck that the AC/DC Adapter is not faulty.Try to operate the Control Module by using batteries instead of the AC/DC Adapter. If

Page 13 - GO NEBULIZER

16WARNINGS:• Only use the Aeroneb® Go nebulizer with components specied in the Instructions for Use. Use of the Aeroneb® Go nebulizer with component

Page 14

17SPECIFICATIONSPhysicalAeroneb Go nebulizer unit (assembled): 40 x 105 x 95 mmAeroneb Go Control Module: 70 x 110 x 32 mmAC/DC adapter: 74 (H) x 72 (

Page 15

18PERFORMANCE:Aerosol ow rate: > 0.3 mL/minAverage tested ow rate: 0.4 mL/minParticle Size:As measured with the Andersen Cascade Impactor:- Spec

Page 16

19Representative particle size distribution graph for Albuterol as per EN13544-1 is shown below:Aerodynamic Size (µm)0.1 1 10 100100.050.00.0Cumulat

Page 17

20REPLACEMENT PARTS AND ACCESSORIES:PART REORDER NUMBERAeroneb® Go System AG-AG4200-XX*Medication Cup Cap AG-AG3010Aeroneb® Go Handset AG-AG4300Mouthp

Page 18

21Appendix 1EMC tables: The following Tables are provided in accordance with IEC/ EN 60601-1-2:Guidance and manufacturer’s declaration – electromagnet

Page 19

Recommended separation distances between portable and mobile RF communication equipment and the Aeroneb GoThe Aeroneb Go nebulizer system is intended

Page 20

Guidance and manufacturer’s declaration – electromagnetic immunityThe Aeroneb Go nebulizer system is intended for use in the electromagnetic environme

Page 21

Immunity test IEC/EN 60601 Test levelCompliance level Electromagnetic environment – guidancePower frequency (50/60 Hz)Magnetic eldIEC 61000-4-83 A/m

Page 22 - PERFORMANCE:

Guidance and manufacturer’s declaration – electromagnetic immunityThe Aeroneb Go nebulizer system is intended for use in the electromagnetic environme

Page 23 - Cumulative Mass (%)

Table of ContentsIntroduction ... 1Important Safeguards ... 2W

Page 24 - Go Handset AG-AG4300

This page has been intentionally left blank

Page 26

©2013 Aerogen LimitedAG-AG4120-EN Rev D 2013 Manufacturing no.: 30-553 Rev D Aerogen Limited,Galway Business Park,Dangan,Galway,Ireland.www.aerogen.c

Page 27

This page has been intentionally left blank

Page 28

1INTRODUCTION Be sure to read these instructions fully before use.Thank you for purchasing the Aeroneb® Go Nebulizer. Certain conditions such as asthm

Page 29

2IMPORTANT SAFEGUARDSWhen you use electrical products, always follow basic safety precautions. As with any electrical device take particular care arou

Page 30

3• Do not use components or attachments that are not recommended by Aerogen. Refer to the Replacement Parts and Accessories section in this manual (p

Page 31

4• Do not nebulize alcohol-based drugs in an oxygen-rich/high pressure environment.• In event of loss of mains power, remove the AC/DC cable from th

Page 32

5WARRANTYThe Aeroneb® Go Control Module and AC/DC Adapter are warranted for a period of two years from the date of purchase against defects in manufac